<6. PROCEDURES FOR CONVERTER OPERATION>
6-22
IM 12E04A02-02E
3rd Edition : May. 31, 2010-00
In step 4, if the set value is beyond the permissible range (-10 to 110), "not" is displayed. Reset
the message by pressing the [>] key and repeat the operations from step 2.
(4) Setting Line-segment Function Output (FUNCTION "6")
Analog output can be provided in the form of a line segment. Set the break point for such a line
segment in this function. A line segment is output, being the set value (percentage of the output
range span) as the break point at 50 % of the output signal span (4 to 20 mA or 1 to 5 V). For
details on the function, see Subsection 5.2.2 (1). The following is the procedure for the setting
operations:
(Example of display)
1. Select FUNCTION "6" in the <PROGRAM1>
=> 00050 %
mode.
2. Press the [>] key.
=> 00050 %
The most signi
fi
cant digit
fl
ashes.
3. Select a digit with the [>] key and set it to the
=> 00060 %
desired value using the [^] or [v] key.
The
fl
ashing digit moves to the right
and when the least signi
fi
cant digit is
reached, it returns to the most
signi
fi
cant digit.
4. Press the [ENT] key to enter the set value.
=> 00060 %
The entire display
fl
ashes once.
In step 4, if the set value is beyond the permissible range (0 to 100 %), "not" is displayed. Reset
the message by pressing the [>] key and repeat the operations from step 2.
(5) Setting Change-over Point in Auto-ranging (FUNCTION "7")
If auto-range is used, the range is changed over at the point set here. The change-over point is
displayed and set at a ratio (%) to the upper limit setpoint of the range. The following shows the
operating manner.
• The time to change over from the lower range to the higher range is the point where
measured values become large than the change-over point in the lower range.
• The timing to change over from the higher range to the lower range is a point where
measured values become 10 % smaller than the change-over point in the lower range.
The following describes the setting operations.
(Example of display)
1. Select FUNCTION "7" in the <PROGRAM1>
=> 00080 %
mode.
2. Press the [>] key.
=> 00080 %
The most signi
fi
cant digit
fl
ashes.
3. Select a digit with the [>] key and set it to the
=> 00090 %
desired value using the [^] or [v] key.
The
fl
ashing digit moves to the right
and when the least signi
fi
cant digit is
reached, it returns to the most
signi
fi
cant digit.
4. Press the [ENT] key to enter the set value.
=> 00090 %
The entire display
fl
ashes once.
In step 4, if the set value is beyond the permissible range (70 to 100 %), "not" is displayed. Reset
the message by pressing the [>] key and repeat the operations from step 2.