9-5
IM CW140-E
Electric Energy Measure Mode
9
9.3 Integration in Electric Energy Measure Mode
●
Start of Integration
Even if you have finished setting the time and commanded the start of
integration by an external trigger, integration does not begin unless you press
the
key (status LED indicator comes on). Integration begins only when
setting of the time has been completed, the START&STOP key has been
pressed, and the
mark comes on.
1
Indicates the status in which the starting time of integration is set and
that the CW140 is ready for integration.
2
Indicates the status in which the START&STOP key is pressed and that
the LED indicator comes on.
3
Indicates that the START&STOP key should be pressed one more time
to begin integration.
Status LED
indicator
START&STOP key
CW140
main unit
1
Location of START&STOP key
and LED indicator
2
3
LED indicator
comes on.
LED indicator
goes out.
LED indicator
comes on.
Ready for
integration
(STANDBY)
Ready for
integration
(STANDBY)
Integration in
progress
Press the START&STOP
key to begin integration.
Scheduled starting
time of integration
TIP
• The changing of settings is not allowed during integration (viewing only).
• The setting of a time that precedes the current time is not allowed. In addition, the CW140 shows an
error message if, after time setting, the preset starting time of integration has expired before integra-
tion begins.
●
End of Integration
If the end of integration has been specified by setting the time or using a timer,
the CW140 automatically stops integration and the
mark
appears when the preset time is up.
• Forcibly stopping integration before the preset time is up
Press the START&STOP key.
The following message appears.
Forcibly stops
Integration.
Scheduled ending
time of Integration
Integration in
progress
Stop integrate?
ENTER: YES ESC: NO
ESC
To stop
Integration
, press the key.
To cancel the procedure for stopping
Integration
, press the key.
Содержание CW140
Страница 1: ...User s Manual CW140 CLAMP ON POWER METER IM CW140 E IM CW140 E 3rd Edition July 2001 YK ...
Страница 2: ......
Страница 14: ......
Страница 72: ......
Страница 102: ......
Страница 138: ......
Страница 152: ......
Страница 160: ......
Страница 200: ......
Страница 256: ......
Страница 274: ......
Страница 276: ......
Страница 286: ......
Страница 289: ......