IM CW140-E
18-8
Communication Execution Errors (200-299)
Error Message
Error Code
Description and Corrective Measures
Execution error
Setting conflict
Data out of range
Too much data
Illegal parameter value
Hardware missing
Cannot execute the command.
Check the relevant setting parameters.
Check the setting range.
Check the data byte length.
Check the setting range.
Check the availability of options.
200
221
222
223
224
241
Communication Query Errors (400-499)
Error Message
Error Code
Description and Corrective Measures
Query INTERRUPTED
Query UNTERMINATED
Query DEADLOCKED
Query UNTRMINATED after indefinite
response
Check the transmission and reception order.
Check the transmission and reception order.
Limit the length of the program message including the
<PMT> to 1024 bytes or less.
Do not enter queries after *IDN?.
410
420
430
440
System Error (912)
Error Message
Error Code
Description and Corrective Measures
Fatal error Communication-driver
Needs servicing.
912
Other (350 and 390)
Error Message
Error Code
Description and Corrective Measures
Queue overflow
Overrun error (RS-232C)
Read the error queue.
Use a lower baud rate.
350
390
18.2 Messages and Corrective Measures
Содержание CW140
Страница 1: ...User s Manual CW140 CLAMP ON POWER METER IM CW140 E IM CW140 E 3rd Edition July 2001 YK ...
Страница 2: ......
Страница 14: ......
Страница 72: ......
Страница 102: ......
Страница 138: ......
Страница 152: ......
Страница 160: ......
Страница 200: ......
Страница 256: ......
Страница 274: ......
Страница 276: ......
Страница 286: ......
Страница 289: ......