19-17
IM CW140-E
Specifications
19
19.2 Specifications of Current Clamps
(3)
Specifications of Clamp-on Probe (96032)
Model No.
960 32
(Clamp B)
Measurement range
0-1000 Arms AC (1414 Apk) 5 minutes
Output voltage
0-0.25 Vrms AC (0.25 mV/A)
Accuracy
Amplitude
±
1.0%rdg
±
0.2 mV (45 Hz to 66 Hz)
Phase
±
1.0
°
(50 A or more, 45 Hz to 66 Hz)
(for temperature of 23
°
C
±
5
°
C, relative humidity of 35-75%, and sinewave input)
Temperature coefficient
±
0.05%/
°
C in ranges of 5-40
°
C
Maximum allowed current
1000 Arms AC : 5 minutes, 700 Arms AC :Continuous (45 Hz to 66 Hz)
Output impedance
Approximately 100
Ω
(max.)
External magnetic field effects 0.5 A equivalent or less (at 400 A/m, 50/60 Hz)
Connector position effects
±
0.5% (at 200-1000 A, 45 Hz to 66 Hz)
Used circuit voltage
600 Vrms AC maximum
Withstand voltage
2.2 kVrms AC for one minute
(across core and casing, and across core and output)
Measurable connector diameter Ø65 mm maximum,
Bus bar: 65
⫻
70 mm max.
Operating temperature
5 to 40
°
C, 35 to 80% RH or less (no condensation)
and humidity ranges
Storage temperature
-20 to 60
°
C, 90% RH or less (no condensation)
and humidity ranges
Operating altitude
2000 m max. above sea level
External dimensions
Approx. 100 (W)
⫻
172.5 (H)
⫻
32 (D) mm
Weight
Approx. 500g
Output cable length
Approx. 3 meters
Accessory
Instruction manual (1)
Ring markers (4 colors
⫻
2), L4007MG
Содержание CW140
Страница 1: ...User s Manual CW140 CLAMP ON POWER METER IM CW140 E IM CW140 E 3rd Edition July 2001 YK ...
Страница 2: ......
Страница 14: ......
Страница 72: ......
Страница 102: ......
Страница 138: ......
Страница 152: ......
Страница 160: ......
Страница 200: ......
Страница 256: ......
Страница 274: ......
Страница 276: ......
Страница 286: ......
Страница 289: ......