14-9
IM CW140-E
External I/O Functions
14
14.3 Event Input
When continuously measuring such data items as electric energy, you can
introduce a signal indicating if a load is active or inactive. It is therefore
possible to manage data that informs you of the relationship between electric
energy and the way the load is being operated.
●
Input Terminals (LOGIC INPUT)
For details on connection, refer to "• Connection of External I/O Terminals" in
Section 2.3.
●
Input Circuit
Same as the circuit shown in Section 14.2, "External Control Input."
●
Input Levels
Same as the level shown in Section 14.2, "External Control Input."
TIP
• To enable event input, set the Event Input option of the SAVE/PRINT ITEM screen to ON.
• The H/L level data of event input is saved or printed at preset output intervals.