IM CW140-E
7-14
7.4
Sampling Frequencies and Integration
Periods
This section describes periods for which the sampling and integrating of measured values are
performed.
●
Sampling in Instant, Electric Energy, or Demand Mode
Sampling period: Approx. 10 kHz
Integration period: 120 msec
50 Hz
100 msec
20 msec
Frequency
measurement
input
Zero crossing
signal
V1
(Sine wave)
The integration period synchronizes with the zero crossing signal of the
frequency measurement input and is defined as a time period which begins
with fixed time (100 ms) and ends with the next leading edge of the zero
crossing signal. The voltage and current RMS, power, and reactive power
(when only the reactive power method is used) values are calculated during
the integration period.
Example:
Frequency (Hz)
Integration period (msec)
Number of sine waves
50
120
6
60
116.67
7
400
102.5
41
1K
101
101