Precautions for Safe Measurement
3
3.6 Connecting a Power Supply and Turning It On/Off
●
Recharging the NiMH Battery Pack
The optional dedicated NiMH battery pack is not fully charged when shipped
from the factory. Before use, recharge the battery pack to its full level. When
recharging it, use the AC adapter.
WARNING
• When recharging the NiMH battery pack, be sure to do so through the CW140 main unit.
• When recharging the NiMH battery pack, keep the ambient temperature within the range from
10°C to 35°C. Recharging the battery pack outside this range may result in an insufficient
amount of charge, solution leakage, or battery heating.
●
Procedure for Recharging the NiMH Battery Pack
Follow the steps below to recharge the NiMH battery pack.
1. With the battery pack installed as instructed earlier, connect an AC
adapter to the CW140.
2. At this point, keep the power switch on the CW140 turned off. The LED
indicator beside the AC adapter jack comes on, indicating that the battery
pack is being recharged. When recharging is complete, the LED indicator
flashes rapidly.
NOTE
If the power switch on the CW140 is turned on, the NiMH battery pack is not recharged. In that
case, the CW140 is powered from the AC adapter.