7-7
IM CW140-E
Common Functions of All Modes
7
Press the
key and select the desired current range from AUTO, and
fixed ranges (varies depending on the clamp type selected).
Pressing the A RANGE key changes the function key labels to allow you to
select the current range as shown below.
20A
50A
100A
200A AUTO
200A 500A 1000A AUTO
50A
100A 200A
500A AUTO
Clamp A
Clamp C
Clamp B
Press the corresponding function key to select the desired range.
NOTE
If continuous measurement (logging, integration, or demand) is performed when in AUTO range,
the range will be fixed to the value when the START&STOP key is pressed.
TIP
• Ranges cannot be changed during continuous measurement.
• It also cannot be changed when in electric energy mode and the elapsed time, “PASSAGE TIME”, is
not at zero, or in demand mode and the remaining time, “DEMAND REST TIME” is not at zero.
• Be sure to press the
key for 3 seconds or more and then clear the integrated value before
changing the ranges.
●
Use of AUTO Range
If measurement is performed when in AUTO range, the range may become
unstable due to the input of irregular pulse waveforms. In this case, change
the range to a fixed value and restart measurement.
When in AUTO range measurement, voltage and current ranges change,
which may result in different ranges for the same power or measured value.
7.3 Ranges and Number of Digits
Содержание CW140
Страница 1: ...User s Manual CW140 CLAMP ON POWER METER IM CW140 E IM CW140 E 3rd Edition July 2001 YK ...
Страница 2: ......
Страница 14: ......
Страница 72: ......
Страница 102: ......
Страница 138: ......
Страница 152: ......
Страница 160: ......
Страница 200: ......
Страница 256: ......
Страница 274: ......
Страница 276: ......
Страница 286: ......
Страница 289: ......