< 4. Piping >
4-8
IM 11M12D01-01E
4.2 Piping for a System to Perform Automatic Calibration
The piping for a system to perform automatic calibration is shown in Figure 4.10.
Averaging Converter (AV550G)
Heater
Signal
Contact
output
Calibration
contact outputs
Span gas cylinder
(Instrument air)
Zero gas cylinder
Flowmeter
Solenoid valve
Needle valve
Air set
Pressure reducing valve
Needle valve
Flowmeter
Reference
gas line
Calibration gas line
Solenoid valve
Detector
Power supply
Instrument air
Contact input
Contact output
Analog outputs:
Averaged and
individual outputs
Stop valve
F4-10E.ai
Figure 4.10 Typical Piping for a System to perform Automatic Calibration
The following outlines some points to note regarding the piping for this system.
• Install a solenoid valve through a nipple on the calibration gas inlet of the detector.
CAUTION
The solenoid valve should be connected directly to the detector. If any piping is present between
the detector and the solenoid valve, condensation may develop in the pipe, causing damage to the
detector by rapid cooling when the calibration gas is introduced.
• When a high temperature detector is used and the pressure of a sample gas is negative, connect
an ejector assembly on the sample gas outlet of the high temperature probe adapter. (See
Figure 4.3 in Subsection 4.1.4.)
• When a high temperature detector is used and the pressure of a sample gas is 0.49 kPa or
higher, it is recommended that a throttle (e.g., needle valve) be installed on the sample gas outlet
of the high temperature probe adapter. (See Figure 4.4 in Subsection 4.1.4.)