< 12. Troubleshooting >
12-13
IM 11M12D01-01E
(3) Conditions for releasing alarm status
(a) If calibration results show improvement compared with previous results e.g. higher cell e.m.f.,
then it is assumed that clogging has been cleaned away, or the cell replaced. Alarm status
and previous data are cleared (but calibration history is not cleared).
(b) If you clear the calibration history, then alarm status and past calibration data is also cleared.
(c) If you disable alarm functions, then alarm status is cleared (but calibration history is not
cleared).
(4) What to do when an alarm occurs
It is possible that the detector tip filter is clogged. Refer to Chapter 11 Inspection and
Maintenance ( on Maintenance for cleaning procedure ).
(5) Alarm setting
(a) On the basic panel display, touch the Setup key to display the [Execution / Setup] display.
(b) Select [Setup] to display the Setup screen.
(c) Select [Other] the [Asymmetry alarm]. The screen below is displayed.
(d) From the [Enable/Disable] window, select [Enable] and touch the Enter key.
Enter
F12-13E.ai
Asymmetry alarm
Asymmetry alarm
Disable
Enable
See Manual for details !
Inside temp. alarm log:
ON 2 4 Jun. 2 0 0 4 16 : 25
OFF 2 4 Jun. 2 0 0 4 16 : 25
ON 2 4 Jun. 2 0 0 4 16 : 23
OFF 2 4 Jun. 2 0 0 4 16 : 25
ON 2 4 Jun. 2 0 0 4 16 : 22
OFF 2 4 Jun. 2 0 0 4 16 : 23
F12-14E.ai
Ch1
Figure 12.13 Asymmetry Alarm Screen
Figure 12.14 Inside Temperature Alarm Log
NOTE
For these alarm functions to be effective, first enable them, then calibrate three times at intervals of at
least 5 days. After that the function will be effective.
9) Alarm 9: Inside temperature alarm
Using the averaging converter in the ambient temperature of more than 50°C may cause a failure in
internal electronic circuit. An IC temperature sensor is installed inside the control card of the averaging
converter and monitors the internal temperature constantly.
If the internal temperature exceeds the limit, an inside temperature alarm will be generated. The
limit is approximately 50°C in terms of case ambient temperature. The occurrence/cancellation time
of this alarm is recorded for the last three times. The record can be viewed in the [Detailed data]
display (Figure 12.14). For frequent occurrence of this alarm, take a measure to lower the ambient
temperature. By assigning the inside temperature alarm to the contact output, a contact may be used
as an ON/OFF switch for cooling equipment (e.g., fan).