< 7. Startup >
7-3
IM 11M12D01-01E
[Disconnect power to the heater; this will generate an error ]
If you disconnect the signal wiring from the detector then this will generate an error, but this allows you
to change the detector model settings without detector warm up.
1) Remove all thermocouple (TC) signal wiring from channel cards which connect to ZO21D
detectors.
2) Apply power to the averaging converter.
3) Error messages are displayed for these channels, but you can set their parameters.
4) Touch the Setup key, and the [Execution/Setup] display appears.
5) Select [Commissioning] --> [Basic setup] --> [Detector], then the Detector Selection display
(Figure 7.2) appears.
6) Select channels to be connected to ZO21D, and change [ZR22G] to read [ZO21D].
7) After changing the channels settings of all channels to be connected to ZO21D detectors, turn off
converter power.
8) Reconnect the removed thermocouple (TC) signal wiring to the channel cards.