< 5. Wiring >
5-10
IM 11M12D01-01E
5.2.8 Wiring for Solenoid Valve for Automatic Calibration
This wiring is for operating the solenoid valve to switch zero and span gases during calibration in the
automatic calibration system and for operating the solenoid valve installed on the calibration gas line
to the detector of each channel. The contact output for solenoid valve is also used for operating the
solenoid valve on the blow back piping. For the wiring and piping of the system, refer to Section 5.3,
Wiring and Piping Examples.
Use shielded cables that withstand temperatures of at least 80°C for the wiring for solenoid valves.
The shields should be connected to cable shield ground terminals on the case of the averaging
converter.
If a cable gland is not installed on the wiring hole of the averaging converter, use a wire with an outside
diameter of 17 mm or smaller. If installed, use a wire with an outside diameter of 6 to 12 mm. Like the
signal wiring to detectors, the shields should be connected to cable shield ground terminals on the
case of the averaging converter.
All contact outputs for solenoid valves are voltage free, dry contacts (mechanical relay contact
outputs). The contacts are open under normal conditions (when deenergized).
The contact rating is 250 V AC, 1A or 30 V DC, 1A.
If optional 24 V outputs are specified for solenoid valves (Option Code “/24”), no external power
supply for solenoid valves is required. The solenoid valves are powered from the AV550G Averaging
Converter. The maximum current that can be supplied to the solenoid valve is 50 mA. Use solenoid
valves that consume not more than 1.2 W to operate.
26 27 28 29
69 70 71 72
44 45 46 47 48 49
89 90 91 92
Averaging Converter
Shield
Basic Power Unit
Expansion Power Unit
Solenoid Valve for Switching
Zero/Span Gases
Power Supply to Solenoid Valve
Solenoid Valve for Calibration Gas
to Detector CH1
Power Supply to Solenoid Valve
Solenoid Valve for Calibration Gas
to Detector CH2
Power Supply to Solenoid Valve
: If optional 24 V outputs are specified for solenoid valves (Option Code "/24"), no external power
supply for solenoid valves is required. The solenoid valves are powered from the AV550G
Averaging Converter. Never connect external power sources in the wiring for solenoid valves.
*
*
*
*
F5.-12E.ai
Figure 5.12 Wiring for Solenoid Valve for Automatic Calibration