< 2. Specifications >
2-2
IM 11M12D01-01E
±5% Maximum value of set range ;
0 to 50 vol% O
2
or more and up to 0 to 100 vol% O
2
range
(Sample gas pressure: within ±0.49 kPa)
Drift:
(Excluding the first two weeks in use)
(Excluding the case where the reference gas is by natural convection.)
Both zero and span ±2% Maximum value of set range/month
Response Time :
Response of 90% within 5 seconds. (Measured after gas is introduced
from calibration gas inlet and analog output start changing.)
2.2 General-purpose Separate type Detector and
Related Equipment
General-purpose separate type detector ZR22G can be used in combination with the probe protector
ZO21R (see Subsection 2.2.2).
2.2.1 ZR22G General-purpose Separate type Detector
Sample Gas Temperature:0 to 700 °C (Probe only)
It is necessary to mount the cell using Inconel cell-bolts when the
temperature is greater than 600 °C.
700 to 1400 °C (with High Temperature Probe Adapter)
For high temperature sample gas, apply 0.15 m long probe and High
Temperature Probe Adapter ZO21P.
Sample Gas Pressure: -5 to +250 kPa (When the pressure in the furnace exceeds 3 kPa,
it is recommended that you compensate the pressure. When the
pressure in the furnace exceeds 5 kPa, you must perform pressure
compensation.) For 0.15 m probe, 0.5 to 5 kPa. No pressure fluctuation
in the furnace should be allowed.
Note:
When the detector is used in conjunction with a check valve and a ZA8F Flow Setting Unit,
the maximum pressure of sample gas is 150 kPa. When with a check valve and a ZR40H
Automatic Calibration Unit, it is 200 kPa. If the pressure of your sample gas exceeds these
limits, consult with Yokogawa.
Probe Length:
0.15, 0.4, 0.7, 1.0, 1.5, 2.0, 2.5, 3.0, 3.6, 4.2, 4.8, 5.4 m
Probe Material: SUS 316 (JIS)
Ambient Temperature: -20 to +150 °C
Reference Gas System: Natural Convection, Instrument Air, Pressure Compensation
(other than for probe length 0.15 m)
Instrument Air System (excluding Natural Convection) :
Pressure; 200 kPa + the pressure inside the furnace (It is recommended
to use air which has been dehumidified by cooling to dew point -20
°C or less, and dust or oil mist removed.)
Consumption; Approx. 1 Nl/min
Material in Contact with Gas: SUS 316 (JIS), Zirconia, SUS 304 (JIS) (flange), Hastelloy B,
(Inconel 600, 601)
Construction:
Heater and thermocouple replaceable construction.
Non-explosion-proof JIS C0920 / equivalent to IP44D.
Equivalent to NEMA 4X/IP66 (Achieved when the cable entry is
completely sealed with a cable gland in the recirculation pressure
compensated version.)