< 10. Other Functions >
10-8
IM 11M12D01-01E
Entering the sampling interval
4) On the basic panel display, touch the Setup key to display the [Execution/Setup] display.
5) Select [Maintenance] then [Display setup] and touch the Enter key.
6) Select [Trend graph] and tough the Enter key.
7) Select [Sample interval] and touch the Enter key and the numerical entry display appears. You
can enter a value for the sampling interval in the range 1 to 30 seconds.
Setting trend graph vertical scale
8) From the [Trend graph] display, select [Upper limit] and touch the Enter key. The numerical entry
display appears, and you can enter the upper limit value. The lower limit value can be entered in
a similar manner. Both upper and lower limit values can be entered in the range 0 to 100% O
2
,
however the upper limit value must be set higher than the lower limit value by at least 1% O
2
.
<Default setting>
The graph parameters are set to the following default prior factory shipment or after data initialization.
Table 10.2
Item
Default
Displayed channels
All channels
Sampling interval
30 seconds
Upper limit
25% O
2
Lower limit
0% O
2