< 10. Other Functions >
10-15
IM 11M12D01-01E
10.4.5 Running an Indication Check
Manual Indication Check
A manual indication check is performed by following prompts on the screen, and manually opening
or closing valves of calibration gas or third check gas. When solenoid valves are used, valves
are opened and closed automatically at the same time that messages appear on the screen. The
procedure for indication check is virtually the same as for calibration. Taking a 3-point indication check
using zero, span, and third gases for example, the procedure is described below.
<Procedure>
1) Confirm that zero, span, and third gas flow control valves are fully closed, then adjust supply gas
pressure to a sample gas pr50 kPa (but a maximum of 300 kPa) using regulator valves
or air sets on the zero, span and third check gas cylinders.
2) Touch Setup key on the basic panel display to switch to the [Execution/Setup] display.
3) Touch [Indication check] and the Indication check display appears.
4) Touch [Manual Ind. check] then the Enter key. The Manual Ind. check Channel select display
shown in Figure 10.13 appears.
5) Select the channels to be checked (channels in warm-up or error status cannot be selected).
Here Channel 2 is taken as an example.
6) Select [Span gas ind. check] and touch the Enter key. A display like Figure 10.14 appears.
Enter
Manual ind. check
Ch1
Ch2
Ch3
Ch4
Ch5
Ch6
Ch7
Ch8
F10-13E.ai
Select ind. check channel
Enter
Zero gas ind. check
Third gas ind. check
F10-14E.ai
Manual ind. check
Span gas ind. check
Ch2
Figure 10.13 Channel select screen Figure 10.14 Manual Span Ind. check
7) Open the span gas flow control valve, and adjust flow rate to 600 ± 60 ml/min. If auto indication
check piping and wiring are installed, then span gas starts to flow automatically.
8) Select [Start ind. check] touch the Enter key (Figure 10.15). An oxygen concentration trend
graph like Figure 10.16 is displayed, and a [CHK.TIME] status mark blinks.
Watch the trend graph and confirm that the measured value stabilizes. Compare the measure
span gas concentration displayed at the bottom of the display against the certified concentration
of span gas to check for any discrepancy or error.
Enter
F10-15E.ai
Ch2
Manual ind. check
Open span gas valve.
Set flow span gas to
600ml/min.
Start ind. check
Cancel ind. check
F10-16E.ai
Enter
21.00%
Ch2
1min./div.
0.98%
20.9 %O2
CHK. TIME
Manual ind. check
Channel No.
under calibration
Setting value of
span gas concentration
Setting value of
zero gas concentration
Actual measurement
value
Figure 10.15 Manual Indication Check Start
Figure 10.16 Span Indication Check