< 7. Startup >
7-2
IM 11M12D01-01E
7.3 Set Valve Type
Set valves (which depend on your system configuration) as follows:
1) If there is a shut-off valve at the calibration gas inlet, shut it.
2) Assuming that instrumentation air supply is used as the reference gas, adjust air set secondary
pressure to approximately 50 kPa above that of the sample gas -- if there is a check valve, then
adjust it to approximately 150 kPa above that of the sample gas (but pressure should be no
greater than 300 kPa). Adjust reference gas flow so that it is in the range 800~1000 ml/min.
7.4 Setting Detector Model
The default detector model set before shipment from the factory is ZR22G. If you plan to use the
ZO21D then before starting warmup you need to change the default setting. If you do not change the
setting before starting warmup then detector temperature control will not work properly and serious
damage to the detector may result.
There are two ways of changing this setting before warmup, as follows:
[ How to use Channel-card “Hot Swap” feature ]
This converter allows you to toggle the power supply of individual channel cards OFF/ ON while power
to the converter is applied. This method also allows you change the detector model settings without
detector warm up.
1) Touch the Setup key, and the [Execution/Setup] display appears.
2) Use the pointer cursor key to move the pointer to [Maintenance] and touch the Enter key.
3) On the Maintenance display, select [Channel card power] and touch the Enter key. A display like
that shown in Figure 7.3 appears.
4) On the [Channel card power] display, select the channel no. of a channel connected to a ZO21D
then touch the Enter key.
5) A [Disable/Enable] selection window is displayed. Select [Disable] and touch the Enter key.
Confirm that the status display lamp of the corresponding channel turns off.
6) Repeat the above procedure for each channel connected to a ZO21D and disable power.
7) Touch the Home key to return to the [Execution/Setup] display.
8) Select [Commissioning] --> [Basic setup] --> [Detector], then the Detector Selection display
appears.
9) Select each channel to be connected to a ZO21D and change the ZR22G default to ZO21D.
10) After changing the settings for all channels to be connected to ZO21D detectors, turn off
converter power.
Ch2 : ZO21D
Ch3 : ZR22G
Ch4 : ZO21D
Ch5 : ZR22G
Ch6 : ZO21D
Ch7 : ZR22G
Ch8 : ZO21D
Detector
Enter
F7-2E.ai
Ch1 : ZR22G
ZR22G
Enter
Ch1 : ZR22G
Ch2 : Enable
Ch3 : Enable
Ch4 : Enable
Ch5 : Enable
Ch6 : Enable
Ch7 : Enable
Ch8 : Enable
F7-3E.ai
Disable
Channel card power
Figure 7.2 Detector Selection
Figure 7.3 Channel Card Power Setting