<8. Detailed Data Setting>
8-12
IM 11M12D01-01E
< Setting Procedure >
1) Touch the Setup key, and the [Execution/Setup] display appears.
2) From [Commissioning] select [Contact setup] and touch the Enter key.
3) Select [Output contact no.] to set up and touch the Enter key.
4) Select [Alarms] or [Others] and touch the Enter key. You can set the condition(s) allocated to
contact output. (Figure 8.14).
5) Select ON/OFF pulldown to set items ON and allocate the corresponding condition to contact
output. You can allocate multiple conditions to a contact output, then if any of the conditions is
true there will be contact output.
Enter
Low Low alarm:
Low alarm: ON
High alarm: OFF
High High alarm: OFF
Cal alarm: OFF
Cal gas low: OFF
Proc. gas alarm: OFF
Error: OFF
Internal temp: OFF
F8-14E.ai
Contact1
Alarms
OFF
Figure 8.14 Alarm Function Allocation
<Default setting>
Contact output setting are set to defaults in Table 8.5 prior factory shipment or after data initialization.
Table 8.5 Contact function initial defaults
Item
Contact
output 1
Contact
output 2
Contact
output 3
Contact
output 4
Alarms
High High Alarm
High Alarm
ON
Low Alarm
ON
Low Low Alarm
CAL Coefficient Alarm
CAL Gas Pressure Low
Process Gas Alarm
Temperature Alarm
Error
Other statuses
(Under) warmup
ON
Output Range Switching
(Under) Calibration
ON
(Under) Indication check
(Under) Maintenance
Blow back
3rd Check Gas
Normally Closed/Open
N.O.
N.O.
N.O.
N.O.