< 6. Components >
6-2
IM 11M12D01-01E
6.1.2 High Temperature Detector (Model ZR22G-015)
Sample gas outlet
When a negative measurement gas
pressure is used, connect the auxiliary
ejector assembly.
When the measurement gas is
high temperature and high-pressure,
and does not fall below 700°C,
connect a pressure control valve
(e.g. a needle valve).
Separate type High temperature Detector (ZR22G-015)
When the temperature of the measurement gas is
between 700°C and 1400°C, mount this detector
with a ZO21P high temperature probe adapter.
F6-2E.ai
Flange
Selectable from JIS standards,
ANSI standards or DIN standards
High Temperature Probe Adapter (ZO21P)
The probe is made of either SUS 310S or silicon carbide (SiC).
Its length is 0.5, 0.6, 0.7, 0.8, 0.9, 1.0 and 1.5 m.
When using an SiC probe, mount it vertically downward.
Figure 6.2 High Temperature Detector