< 7. Startup >
7-11
IM 11M12D01-01E
5) Select the Analog output to perform Loop check on, and touch the Enter key.
6) The numerical entry display appears. Set output current somewhere in the range 4 to 20 mA.
Touching the Entry key outputs this current value.
7) Touching the Reject key (door with arrow) from the numerical entry display returns you to the
“mA-output loop check” channel-select display.
8) You can then perform a loop check on other analog outputs in the same manner.
7.11 Checking Operation of Contact Input, Contact
Output
After wiring the cables you can perform a contact I/O check to display the status of contact inputs and
to check the operation of contact outputs including contact outputs to operate solenoid valves during
automatic calibration.
7.11.1 Contact Output Operation Check
1) Touch the Setup key to switch to the [Execution/Setup] display.
2) Select [Maintenance] and touch the Enter key.
3) From the Maintenance display select [Contact check], and touch the Enter key.
4) The Contact check display shown in Figure 7.18 allows you to select contacts to check.
To check the individual error contacts for channels 1 to 8, select [Output contact ch(Channel
contact output)] or to select Function contacts (contacts 1 to 5) select [Output contact
ctrl(Common contact outputs)].
5) Next the Contact no. select display is displayed. Select the contact you want to check, then touch
the Enter key, and the [OFF] / [ON] selection window opens. Select then touch the Enter key to
open or close the corresponding contact output.
F7-18E.ai
Enter
Output contact ctrl
Solenoid valve contacts
Input contacts
Contact check
Output contact ch
F7-19E.ai
Enter
Ch1 :
Ch2 : ON
Ch3 : OFF
Ch4 : OFF
Ch5 : OFF
Ch6 : OFF
Ch7 : OFF
Ch8 : OFF
Output contact ch
OFF
Figure 7.18 Contact check display
Figure 7.19 Contact select display
CAUTION
1) Individual error contacts do not operate if corresponding channel is not installed.
2) If you perform a Channel contact output Open / Close check then Error2 is generated. The
reason is that Channel contact output is used as Sensor heater power safety switch, and this
turns off power to the heater. If this error occurs, to reset it remove the channel card then reinsert
it (hot swap functions), or turn power OFF then back ON.