10.4 Indication Check
In addition to zero and span calibration gases, you can select a third check gas for a 3- point check.
This is another check to determine if there is any discrepancy between the measured value and the
certified concentration of the gas.
10.4.1 Mode
There are three modes of indication check operation: manual, semi-automatic, and automatic.
In Manual mode, zero, span, and third gas indication checks can be performed sequentially and
manually. In Semi-auto mode, an indication check can be started by key operation on the display or by
a contact input signal, and then sequentially performed at a preset check time and hold time. In Auto
mode, indication checks are automatically performed at preset intervals.
The following restrictions apply:
· When [Manual] is selected
In this mode, only Manual Indication Check can be performed. (Contact input used to start Semi-Auto
Indication Check has no effect. Auto Indication Check does not start at Auto Indication Check start
time).
· When [Semi-Auto] is selected.
In this mode, Manual Indication Check or Semi-Auto Indication Check can be performed.
(Auto Indication Check does not start at Auto Indication Check).
· When [Auto] is selected.
Indication Check can be performed in any of the three modes.
CAUTION
When [Auto] Calibration mode is selected, you cannot also select [Auto] Indication Check mode. The
reverse also applies: if you select [Auto] Indication Check mode then [Auto] Calibration mode cannot
be selected.
<Setting Procedure>
1) On the basic panel display, touch the Setup key to display the [Execution/Setup] display, then
select [Maintenance].
2) From the [Maintenance] display, select [Indication check setup] then [Mode]. A dropdown menu
opens, allowing you to select between Manual, Semi-auto, or Auto modes (see Figure 10.9).
Enter
F10-9E.ai
Points: Semi-auto-Third
Timing:
Indication check setup
Mode: Semi-auto
Manual
Semi-auto
Auto
Figure 10.9 Indication Check Mode Selection