< 5. Wiring >
5-8
IM 11M12D01-01E
5.2.5 Signal Wiring to Detectors
This wiring is for transmitting signals of cell electromotive force, thermocouple output and cold junction
compensation from a detector to the averaging converter. Use 6-core shielded cables that withstand
temperatures of at least 80°C. The conductor two-way resistance should be 10 Ω or less. Keep the
cables away from the power wiring. If a cable gland is not installed on the wiring hole of the averaging
converter, use a wire with an outside diameter of 17 mm or smaller. If installed, use a wire with an
outside diameter of 6 to 12 mm. The shields should be connected to cable shield ground terminals on
the case of the averaging converter.
1
1
1
1
2
2
2
2
1 2 3 4 5 6
3
3
3
3
4
4
4
4
5
5
5
5
6
6
6
6
To Detector for CH2
ZR22G / ZO21D
CH1
CH2
Detector for CH1
CH3
CH4
Shield
Averaging Converter
Channel Card
CELL
TC
CJ
F5-9E.ai
+
–
+
–
+
–
Figure 5.9 Signal Wiring to Detectors
If the ambient temperature at the detector installation site exceeds 75°C, install a terminal box
and connect to the detector using 600 V silicon rubber insulated glass braided wires. The distance
between the detector and the terminal box should be minimized.
F5-10E.ai
To Detector for CH2
ZR22G / ZO21D
CH1
CH2
Detector for CH1
CH3
CH4
Shield
Averaging Converter
Terminal Box
Silicon Rubber Insulated Glass Braided Wire
Channel Card
CELL
TC
CJ
+
–
+
–
+
–
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
Figure 5.10 Signal Wiring to Detectors (Using Terminal Box)
Содержание av550g
Страница 17: ...Blank Page ...
Страница 45: ...Blank Page ...
Страница 71: ...Blank Page ...
Страница 203: ......
Страница 205: ......
Страница 207: ......
Страница 209: ......
Страница 211: ......
Страница 213: ...Blank Page ...