< 7. Startup >
7-1
IM 11M12D01-01E
7. Startup
This Chapter explains the startup procedures for the Averaging Converter: supplying power, setting
parameters as required, and performing manual calibration.
7.1 Startup Procedure
The startup procedure is as follows:
CAUTION
If you connect Model ZO21D detectors, then you need to change the detector parameters.
Before connecting power, refer to: Section 7.4 Setting Detector Model.
Check valve type setting
Check piping & wiring
Check settings
F7-1E.ai
If connecting ZO21D
Briefly power on
Set detector to ZO21D
(front panel)
Power OFF
If all detectors are ZR22G
Permanent power wiring
Start set parameters
Warmup, then calibrate
Figure 7.1 Startup Procedure
7.2 Check Piping and Wiring
Check piping and wiring (refer to Chapter 4 for piping, and Chapter 5 for wiring).