D8 Series User’s Guide
Chapter 8: Specifications
158
Calibration
Automatic zero and full scale
Analog Ground to Frame Ground
Maximum
40V
DC Common to Frame Ground
Maximum Potential
40V
Open Thermocouple Detection
Pulse type for upscale break detection
Table 8.9
– Thermocouple Range and Resolution
THERMOCOUPLE
TYPE
RANGE IN
°F
RANGE IN
°C
ACCURACY* AT
25 °C AMBIENT
ACCURACY* AT
0 TO 50 °C AMBIENT
°F
°C
°F
°C
J
-350 to 1,400
-212 to 760
±2.2
±1.2
±3.3
±1.8
K
-450 to 2,500
-268 to 1,371
±2.4
±1.3
±3.8
±2.1
T
-450 to 750
-268 to 399
±2.9
±1.6
±5.8
±3.2
S
0 to 3,200
-18 to 1,760
±5.0
±2.8
±8.8
±4.9
R
0 to 3,210
-18 to 1,766
±5.0
±2.8
±8.8
±4.9
B
150 to 3,200
66 to 1,760
±7.2
±4.0
±22.1
±12.3
E
-328 to 1,448
-200 to 787
±1.8
±1.0
±2.9
±1.6
* True for 10% to 100% of span except type B, which is specified for 800° F to 3200° F.
Table 8.10
– RTD Range and Resolution
RANGE
IN °F
RANGE
IN °C
RESOLUTION
IN °C
MEASUREMENT
TEMPERATURE
IN °C
ACCURACY* AT
25 °C AMBIENT
ACCURACY AT
0 TO 50 °C
AMBIENT
°F
°C
°F
°C
-328.0 to
1150.0
-200.0 to
621.1
0.07
25
0.9
0.5
1.2
0.5
400
2.7
1.5
4.1
2.2
Table 8.11
– Input Resistance for Voltage Inputs
RANGE
INPUT RESISTANCE
0 to 10V
50k
Ω
0 to 5V
40k
Ω