Servicing
Pressure Transducer Calibration
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
6-33
A rough check of the pressure accuracy can be made by obtaining the
current barometric pressure from the local weather station or airport and
comparing it to the pressure reading. However, since these pressures are
usually corrected to sea level, it may be necessary to correct the reading to
local pressure by subtracting 0.027 mm Hg per foot of altitude.
Do not try to calibrate the pressure transducer unless the pressure is known
accurately.
1. Remove the cover.
2. Disconnect the tubing from the pressure transducer and connect a
vacuum pump known to produce a vacuum less than 1 mm Hg.
3. From the Main Menu, press
to scroll to Service > press
>
to scroll to
Pressure Calibration
> and press
.
The Pressure Sensor Cal screen appears.
Note
If Service is not displayed, refer to “
, then return to the beginning of this step.
4. At the Pressure Sensor Cal screen, press
to
select
Zero
.
The Calibrate Pressure Zero screen appears.
5. Wait at least 10 seconds for the zero reading to stabilize, then press
to save the zero pressure value.
6. Disconnect the pump from the pressure transducer.
7. Press
to return to the Pressure Sensor Cal screen.
8. At the Pressure Sensor Cal screen, press
to select
Span
.
The Calibrate Pressure Span screen appears.
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...