Troubleshooting
Troubleshooting Guides
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
5-3
Lamp
Remove one cell and look for
blue light in hole of input
block
Lamp heater
Check lamp temperature.
From the Diagnostics menu
choose Temperatures.
±15 volt power supply
Check ±15 volts. From the
Diagnostics menu choose
Voltages.
Cell A or B noise
excessive
Foreign material in one cell
Clean cell.
Defective detector
Interchange detectors at
Measurement Interface
Board connectors to
determine if detector is
defective.
Cell A and B noise
excessive
Foreign material in
cells
Clean cells
Lamp failure
Check for 1.7 volt peak to
peak waveform at lamp
current check point on Lamp
Power Supply Board
±15 volt power supply
Check ±15 volts. From the
Diagnostics menu choose
Voltages.
Pressure transducer
does not hold
calibration
Pressure transducer
Replace pressure transducer
Output signal
noisy
Recorder
Replace or repair recorder
Sample is varying
Run instrument on stable
ozone source. If quiet, no
malfunction.
Foreign material in cell
Clean cell
Sticky solenoid valve
Replace with known good
solenoid valve.
Slow response
Averaging time
Verify averaging time is set
properly
Contaminated optical bench
Clean bench and then
condition system overnight
Table 5-1.
Troubleshooting - General Guide, continued
Malfunction
Possible Cause
Action
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...