Servicing
Pressure Transducer Replacement
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
6-31
Figure 6-12.
Replacing the Detectors
4. Loosen the Allen screw(s) securing the detector(s) in detector block and
remove the detector(s).
5. Install the new detector(s) by following the above steps in reverse.
6. Calibrate the instrument. Refer to the “Calibration” chapter in this
manual.
Pressure Transducer
Replacement
Use the following procedure to replace the pressure transducer.
Equipment Required:
Pressure transducer assembly
Philips screwdriver
1. Turn instrument OFF, unplug the power cord, and remove the cover.
Clamping Block
Clamping Block Screws (2)
Detectors
Detector Screws (2)
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...