Operation
Service Menu
3-72
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
useful to determine how much ozone is produced at a certain percentage of
lamp intensity. The pressure calibration is visible only when the calibrator is
in service mode. For more information on the service mode, see “Service
Mode” earlier in this chapter.
Note
This adjustment should only be performed by an instrument service
technician.
•
In the Main Menu, choose Service >
Manual O
3
Lamp Setting
.
Pressure Calibration
The Pressure Calibration menu is used to calibrate the pressure sensor to
zero, span, or restore factory default values. The pressure calibration is
visible only when the calibrator is in service mode. For more information on
the service mode, see “Service Mode” earlier in this chapter.
The pressure sensor’s zero counts and span slope are displayed on the menu.
Note
This adjustment should only be performed by an instrument service
technician.
•
In the Main Menu, choose Service >
Pressure Calibration
.
MANUAL O3 LAMP SETTING:
LAMP SETTING: 0.1% ?
©ª
INC/DEC
PRESSURE CALIBRATION
RANGE AVG DIAGS ALARM
PRESSURE SENSOR CAL:
>ZERO 72
SPAN 1.1447
SET DEFAULTS
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...