Operation
Range Menu
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
3-9
•
In the Main Menu, choose Range >
O
3
Range
.
•
Use
and
to scroll through a list of choices
.
•
Press
to save the new range.
Table 3-2.
Standard Ranges
C1, C2, and C3 are custom ranges. The analog outputs are arranged on the
rear panel connector as shown in Figure 3-4. See Table 3-3 for channels and
pin connections. For more information about custom ranges, see “Set
Custom Ranges” below.
ppb
ppm
μ
g/m
3
mg/m
3
50
0.05
100
0.1
100
0.10
200
0.2
200
0.20
400
0.4
500
0.50
1,000
1
1,000
1
2,000
2
2,000
2
5,000
5
5,000
5
10,000
10
C1
C1
C1
C1
C2
C2
C2
C2
C3
C3
C3
C3
O3 RANGE:
CURRENTLY: 500.0
SET TO: 1000.0 ?
©ª
CHANGE VALUE
SAVE VALUE
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...