C-Link Protocol Commands
Record Layout Definition
B-44
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Note
If individual relays have been assigned different logic, the response
would be a 4-digit hexadecimal string with the least significant byte (LSB)
being relay no 1.
For example:
Receive:
relay stat 0x0001 (indicates relay no 1 is set to normally
open logic, all others are normally closed)
Receive:
relay stat 0x0005 (indicates relay no 1 and 3 are set to be
normally open logic, all others are normally closed)
set relay open
set relay open
value
set relay closed
set relay closed
value
These commands set the relay logic to normally open or closed for relay
number
value
, where
value
is the relay between 1 and 16. The example that
follows sets the relay no 1 logic to normally open.
Note
If the command is sent without an appended relay number then all the
relays are assigned the set logic of normally open/closed.
Send:
set relay open 1
Receive:
set relay open 1 ok
Record Layout
Definition
The Erec, Lrec, and Srec Layouts contain the following:
•
A format specifier for parsing ASCII responses
•
A format specifier for parsing binary responses
In addition to these, the Erec Layout contains:
•
A format specifier for producing the front-panel displays
In operation, values are read in using either the ASCII or binary format
specifiers and converted to uniform internal representations (32-bit floats or
32-bit integers). These values are converted into text for display on the
screen using the format specifier for the front-panel display. Normally, the
specifier used to parse a particular datum from the input stream will be
strongly related to the specifier used to display it (e.g., all of the floating
point inputs will be displayed with an 'f' output specifier, and all of the
integer inputs will be displayed with a 'd' specifier).
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...