C-Link Protocol Commands
I/O Configuration
B-42
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Send:
set din 1 3 high
Receive:
set din 1 3 high ok
dout
channel
This command reports the index number and output variable and the active
state assigned to output
channel.
The example that follows reports the input
4 to be assigned an index number 11 corresponding to “general alarm” with
the active state being open.
Send:
dout 4
Receive:
dout 4 11 GEN ALARM open
set dout
channel index state
This command assigns digital output
channel
to be assigned to the action
associated with
index
, and assigns it an active state of
state
(open or closed).
Send:
set dout 4 11 open
Receive:
set dout 4 11 open ok
dtoa
channel
This reports the outputs of the 6 or 12 digital to analog converters,
according to
. The example that follows shows that the DAC 1 is
97.7% full-scale.
Send:
dtoa 1
Receive:
dtoa 1 97.7%
Note
All channel ranges are user definable. If any customization has been
made to the analog output configuration, the default selections may not
apply.
Table B-11.
Default Output Assignment
D to A
Function
1
Voltage Output
2
Voltage Output
3
Voltage Output
4
Voltage Output
5
Voltage Output
6
Voltage Output
7
Current Output
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...