Servicing
Detector Replacement
6-30
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
6. Replace the new solenoid by following the previous steps in reverse
making sure that all connections are tight.
7. Perform a leak test. Refer to the “Solenoid Leaks” procedure in the
“Preventive Maintenance” chapter.
Detector
Replacement
Use the following procedure to replace a detector (
).
Equipment Required:
Detector assembly
Allen wrench, 9/64-inch and 1/16-inch
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect the detector electrical connections (CHA and CHB) from
the measurement interface board.
3. Loosen the two screws securing the detector clamping block to the
optical bench and remove the clamping block and detectors
(
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...