Operation
Instrument Controls Menu
3-16
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Pump
The Pump screen allows the user to toggle the pump on and off.
•
In the Main Menu, choose Instrument Controls >
Pump
.
•
Press
to toggle and set the pump on or off.
Temperature
Compensation
Temperature compensation provides compensation for any changes to the
instrument's output signal due to variations in sample gas temperature. The
Model 49
i
Primary Standard can be operated with or without temperature
compensation.
When temperature compensation is on, the display shows the sample gas
temperature (measured by a thermistor on the Interface board). When
temperature compensation is off, the display shows the standard
temperature of 0.0 °C.
•
In the Main Menu, choose Instrument Controls >
Temperature
Compensation
.
•
Press
to toggle and set the temperature compensation on or off.
CUSTOM LEVEL 1:
CONC SETTING: 450 ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
PUMP:
CURRENTLY: OFF
SET TO: ON ?
TOGGLE VALUE
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...