C-Link Protocol Commands
Alarms
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
B-13
Send:
set alarm bench lamp temp max 58.0
Receive:
set alarm bench lamp temp max 58.0 ok
alarm bench temp min
alarm bench temp max
These commands report the bench temperature minimum and maximum
value current settings. The example that follows reports that the alarm
bench temperature minimum value is 15.0
°
C.
Send:
alarm bench temp min
Receive:
alarm bench temp 15.0
set alarm bench temp min
value
set alarm bench temp max
value
These commands set the bench temperature minimum and maximum value
to
value
, where
value
is a floating-point number representing bench
temperature alarm limits in degrees C. The example that follows sets the
alarm bench temperature maximum value to 35
°
C.
Send:
set alarm bench temp max 35
Receive:
set alarm bench temp max 35 ok
alarm conc o3 min
alarm conc o3 max
These commands report the conc O
3
concentrations alarm minimum and
maximum value current settings. The example that follows reports that the
O
3
concentration minimum value is 5.2 ppb.
Send:
alarm conc o3 min
Receive:
alarm conc o3 min 5.2 ppb
set alarm conc o3 min
value
set alarm conc o3 max
value
These commands set the O
3
concentration alarm minimum and maximum
value to
value
, where
value
is a floating-point representation of the
concentration alarm limits. The example that follows sets the O
3
concentration alarm maximum value to 150.
Send:
set alarm conc o3 max
Receive:
set alarm conc o3 max 150 ok
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...