Operation
Diagnostics Menu
3-54
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Cell A/B O
3
The Cell A/B O
3
screen (read only) displays the current O
3
concentration in
each cell, as well as the O
3
reading displayed in the Run screen (the average
of the two cells).
•
In the Main Menu, choose Diagnostics >
Cell A/B O
3
.
Intensities
The Intensities screen (read only) displays the current intensity in Cell A
and Cell B in Hertz. These intensities are read by detectors A and B,
respectively.
•
In the Main Menu, choose Diagnostics >
Intensities
.
FLOWS:
CELL A 0.608 LPM
CELL B 0.612 LPM
RANGE AVG DIAGS ALARM
O3 PPB: 600
CELL A 598
CELL B 602
RANGE AVG DIAGS ALARM
INTENSITIES:
CELL A 100425 Hz
CELL B 100465 Hz
RANGE AVG DIAGS ALARM