Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
3-39
Digital Input Settings
The Digital Input Settings menu displays a list of the 16 digital inputs
available, and allows the user to select the logic state and instrument
parameter for the relay selected.
Note
The digital inputs must be asserted for at least one second for the
action to be activated.
•
In the Main Menu, choose Instrument Controls > I/O Configuration >
Digital Input Settings
.
•
Use
and
to move the cursor up and down.
•
Press
to select a choice.
Logic State
The Logic State screen is used to change the I/O relay to either normally
open or normally closed. The default state is open, which indicates that a
relay connected between the digital input pin and ground is normally open
NON ALARM STATUS ITEMS:
>NONE
SERVICE
UNITS
ZERO MODE
SPAN MODE
SAMPLE MODE
OZ LEVEL 1
ª
RANGE AVG DIAGS ALARM
OZ LEVEL 2
OZ LEVEL 3
OZ LEVEL 4
OZ LEVEL 5
PURGE MODE
DIGITAL INPUT SETTINGS:
>1 NOP SET BACKGROUND
2 NOP OZONATOR LVL 1
3 NOP OZONATOR LVL 2
4 NOP AOUTS TO ZERO
5 NOP AOUTS TO FS
6 NOP NONE
7 NOP NONE
ª
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...