Servicing
Photometer Board Replacement
6-26
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Figure 6-11.
Replacing the Optical Bench
Photometer Board
Replacement
Use the following procedure to replace the photometer board (
).
Equipment required:
Photometer board
Philips screwdriver
Use the following procedure to replace the photometer board.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Remove the connection cables from the board. Note the locations to
facilitate reconnection.
Photometer Lamp
Lamp Heater
Access Port Cell B IN
Absorption Cell B
Access Port Cell B OUT
Detector B
Access Port Cell A IN
Input Block
Absorption Cell A
Access Port Cell A OUT
Output Block
Detector A
Captive Screws (4)
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...