Operation
Service Menu
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
3-79
•
Press
to save the value.
Analog Input Calibration
The Analog Input Calibration menu is a selection of 8 analog inputs (if the
I/O expansion board option is installed) to calibrate, and allows the user to
select the calibration action zero or span. The analog input calibration is
visible only when the calibrator is in service mode. For more information on
the service mode, see “Service Mode” earlier in the chapter.
Note
This adjustment should only be performed by an instrument service
technician.
•
In the Main Menu, choose Service >
Analog Input Calibration
.
ANALOG OUTPUT CAL: SPAN
CONNECT METER TO OUTPUT!
SELECTED OUTPUT: V1
SET TO: 3697
SAVE VALUE
©ª
INC/DEC
SET OUTPUT TO: 10 V
RANGE AVG DIAGS ALARM
ANALOG INPUT CAL:
>INPUT CHANNEL 1
INPUT CHANNEL 2
INPUT CHANNEL 3
INPUT CHANNEL 4
INPUT CHANNEL 5
INPUT CHANNEL 6
INPUT CHANNEL 7
ª
RANGE AVG DIAGS ALARM
ANALOG INPUT CAL:
>CALIBRATE ZERO
CALIBRATE FULL SCALE
RANGE AVG DIAGS ALARM
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...