C-Link Protocol Commands
Alarms
B-12
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Figure B-1.
Flag Status
Alarms
alarm bench lamp temp min
alarm bench lamp temp max
These commands report the bench lamp temperature minimum and
maximum value current settings. The example that follows reports that the
alarm bench lamp temperature minimum value is 52
°
C.
Send:
alarm bench lamp temp min
Receive:
alarm bench lamp temp 52.0 deg C
set alarm bench lamp temp min
value
set alarm bench lamp temp max
value
These commands set the bench lamp temperature minimum and maximum
value to
value
, where
value
is a floating-point number representing bench
lamp temperature alarm limits in degrees C. The example that follows sets
the alarm bench lamp temperature maximum value to 58
°
C.