Servicing
Analog Input Calibration
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
6-15
The Analog Output Cal line displays Zero
Note
When calibrating the analog output, always calibrate zero first and
then calibrate full scale.
5. Use
until the meter reads the value shown in the Set
Output To line, then press
to save the value.
6. Press
to return to the previous screen.
7. Press
to select Calibrate Full Scale.
8. Use
until the meter reads the value shown in the Set
Output To line, then press
to save the value.
Analog Input
Calibration
Use the following procedures to calibrate the analog inputs after replacing
the optional I/O expansion board. These procedures include selecting
analog input channels, calibrating them to zero volts, and then calibrating
them to full scale using a known voltage source.
Calibrating the Input
Channels to Zero Volts
Use the following procedure to calibrate the input channels to zero volts.
1. From the Main Menu, press
to scroll to Service, press
>
to scroll to Analog Input Calibration
,
and press
.
The Analog Input Cal screen displays.
Note
If Service is not displayed, refer to “
, then return to the beginning of this step.
2. At the Analog Input Cal screen, press
to scroll to a channel, and
press .
3. With the cursor at Calibrate Zero, press
.
The screen displays the input voltage for the selected channel.
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...