Servicing
Analog Output Testing
6-12
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Analog Output Testing
The analog outputs should be tested if the concentration value on the front
panel display disagrees with the analog outputs. To check the analog
outputs, you connect a meter to an analog output channel (voltage or
current) and compare the meter reading with the output value displayed on
the Test Analog Outputs screen.
Equipment Required:
Multimeter
Use the following procedure to test the analog outputs.
1. Connect a meter to the channel to be tested.
shows the
identifies the associated channels.
2. From the Main Menu, press
to scroll to Diagnostics, press
>
to scroll to Test Analog Outputs, and press
.
The Test Analog Outputs screen displays.
3. Press
to scroll to the channel corresponding to the rear panel
terminal pins where the meter is connected, and press
.
The Set Analog Outputs screen displays.
4. Press
to set the output to zero.
The Output Set To line displays Zero.
5. Check that the meter is displaying a zero value. If the meter reading
differs by more than one percent, the analog outputs should be adjusted.
Refer to the “Analog Output Calibration” procedure that follows.
6. Press
to set the output to full scale.
The Output Set To line displays Full Scale.
7. Check that the meter is displaying the full scale value. If the meter
reading differs by more than one percent, the analog outputs should be
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...