Preventive Maintenance
Monitoring Detector Frequencies and Noise
4-4
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Monitoring Detector
Frequencies and
Noise
The Model 49
i
Primary Standard measures intensity ratios and not absolute
values. Therefore, a large range of detector frequencies are acceptable for
proper operation of the instrument. The nominal values are 65 to 120 kHz.
These frequencies can be monitored from the Intensities screen in the
Diagnostics menu.
1. Press
to display the Main Menu.
2. Use
to scroll to Diagnostics, press
>
to
scroll to
Intensities
and press
.
The Intensities screen appears.
Degradation of detector frequencies to below 65 kHz indicates either a dirty
cell or low lamp output. In addition to degrading the measured detector
frequency, dirt in the cells can decompose the ozone and give erroneous
readings. Therefore, the cells should first be cleaned and the frequency
re-measured.
Increasing Lamp Output
If the frequencies are still low, the lamp output can be increased by using the
Lamp Setting screen in the Service mode. If the frequency cannot be set
above 65 kHz, replace the lamp.
1. Press
to display the Main Menu.
2. Use
to scroll to Service, press
>
to scroll to
Lamp Setting
and press
.
The Lamp Setting screen appears.
If the Service Mode is not displayed on the Main Menu, use the
following procedure to display it.
a. At the Main Menu, press
to scroll to Instrument Controls >
press
>
to scroll to
Service Mode
> and press
.
The Service Mode screen appears.
b. Press
to toggle the Service Mode to ON.
c. Press
>
to return to the Main Menu.
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...