System Description
Hardware
7-2
Model 49
i
Primary Standard Instruction Manual
Thermo Fisher Scientific
Figure 7-1.
Hardware Components
Optical Bench with
Photometer Lamp
The optical bench has two airtight chambers that contain the sample and
reference gases with a common photometer lamp at one end and two
individual detectors at the other end.
Detector System
The photo-diode in each detector transmits light intensity information to
the measurement interface board for sample measurement computations.
Flow Transducers
Flow transducers monitor the flow of the sample and reference gases and
transmit the data to the measurement interface board.
Pressure Transducer
The pressure transducer measures the pressure of the sample gas.
Solenoid Assembly
Lamp Heater
Assembly
Photometer Lamp
Pump
Flow Transducer A
Flow Transducer B
Detector Assembly A
Detector Assembly B
Pressure Transducer
Optical Bench
Fan
Sample Valve
Power Supply
I/O Expansion
Board (Optional)
Digital Output Board
Motherboard
Reference Valve
Ozonator
Measurement
Interface Board
Ozonator Board
Photometer Board
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...