Servicing
Photometer Lamp Replacement
Thermo Fisher Scientific
Model 49
i
Primary Standard Instruction Manual
6-27
3. Remove the two screws from the top of the board.
4. Pop off the board from the bottom mounting studs and remove the
board.
5. To install the new photometer board, follow previous steps in reverse.
Photometer Lamp
Replacement
Equipment required:
Photometer lamp
Allen wrench, 7/64-inch and 3/32-inch
Screwdriver - flat - medium size
Use the following procedure to replace the photometer lamp.
1. Turn instrument OFF, unplug the power cord, and remove the cover.
2. Disconnect the photometer lamp cable from the LAMP connector on
the photometer board.
3. Remove the two retaining screws from the rear panel access hole cover
on the rear panel and remove the cover.
4. Slide the insulation off the end of the lamp.
5. Loosen Allen screw on lamp clamp.
6. Loosen two screws holding the clamp to the optical bench.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap must
be worn while handling any internal component.
Содержание 49i
Страница 4: ......
Страница 12: ...Figures xii Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 146: ...Troubleshooting Service Locations 5 22 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...
Страница 203: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 5 Figure 8 2 Bench Mounting ...
Страница 204: ...8 6 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific Figure 8 3 EIA Rack Mounting ...
Страница 205: ...Thermo Fisher Scientific Model 49i Primary Standard Instruction Manual 8 7 Figure 8 4 Retrofit Rack Mounting ...
Страница 206: ...Optional Equipment Mounting Options 8 8 Model 49i Primary Standard Instruction Manual Thermo Fisher Scientific ...