Maintenance
Servicing Plan
8-6
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
Forepump Sogevac
Check oil level
Daily
see manufacturer’s
manual
Check gas ballast valve
Monthly
Replace exhaust filter
Annually
Check anti-suck back valve
Annually
Clean fan cover
Annually
1st oil change
Depending on operation
condition
Replace PFPE oil
Annually or depending
on operation condition
Chiller
Replace fluid bag
Depending on operation
condition
see manufacturer’s
manual
Replace condenser filter
Depending on operation
condition
Clean reservoir
Depending on operation
condition
Table 8-1.
Maintenance procedures
Item
Procedure
Frequency
Procedure location