System Description
Sample Introduction System
3-4
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
Sample Introduction System
The sample introduction system of the iCAP Q MS comprises the
peristaltic pump, the nebulizer, the Peltier-cooled spray chamber and the
torch with the injector.
shows a schematic view of the sample
introduction system.
Peristaltic Pump
A peristaltic pump passes the liquid sample to the nebulizer.
shows the peristaltic pump with sample and drain tubing:
Labeled Components: 1=ICP torch with injector, 2=spray chamber,
3=nebulizer, 4=drain to waste (peristaltic pump), 5=to waste, 6=sample to
nebulizer (peristaltic pump), 7=nebulizer gas
Figure 3-3.
Schematics sample introduction
3
2
1
4
6
7
5
Figure 3-4.
Peristaltic pump at inlet system