Maintenance
General Advice for Maintenance
8-4
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
General Advice for Maintenance
When performing maintenance on the iCAP Q mass spectrometer,
observe the following advice:
•
Accurate results can be obtained only if the system is in good
condition and properly calibrated.
•
Preventive maintenance must commence with installation, and must
continue during the warranty period to maintain the warranty.
Thermo Fisher Scientific offers maintenance and service contracts.
Contact your local Thermo Fisher Scientific office for more
information. Routine and infrequent maintenance procedures are
listed in
•
The operator maintenance procedures described in this manual do
not require removing the instrument housing. Thermo Fisher
Scientific assumes no responsibility and will not be liable for
instrument damage and/or operator injury that might result from
using the instrument without the housing attached. Therefore, a
Thermo Fisher Scientific field service engineer must be called if
removal of the instrument housing is required.
To successfully carry out the procedures listed in this chapter, observe
the following rules:
•
Proceed methodically.
•
Always wear clean, talc-free, and lint-free gloves when handling the
components of the iCAP Q mass spectrometer.
Thermo Fisher Scientific recommends white nitrile clean room
gloves (Fisher Scientific P/N 19-120-2947B, size medium, or P/N
19-120-2947C, size large, or Thermo Scientific P/N 23827-0008,
size medium, or P/N 23827-0009, size large).
•
Always place the components on a clean, lint-free surface.
•
Never overtighten a screw or use excessive force.
•
Dirty tools can contaminate your system. Keep the tools clean and
use them exclusively for maintenance and service work at the mass
spectrometer.
Returning Parts
To protect our employees, we ask you for some special precautions when
returning parts for exchange or repair to the factory. Your signature on
Health and Safety Form
confirms that the returned parts have been
de-contaminated and are free of hazardous materials. See
for further information.