Maintenance
Maintaining the Interface
Thermo Fisher Scientific
iCAP Q Operating Manual (P/N 1288090, Revision B)
8-23
6. Mount the skimmer cone insert again.
7. Turn the tool to secure the position of the insert.
8. Remove the tool.
Insert the Skimmer Cone
❖
To insert the skimmer cone
1. Insert the skimmer cone again using the cone tool.
2. Reassemble the sample cone and the cone holder, see
.
3. Close the plasma door.
Maintaining the Extraction Lens
Depending on the type of applications and on the general operating
conditions, the extraction lens must be cleaned or replaced.
Removing the Extraction Lens
❖
To remove the extraction lens
1. Set the instrument to standby mode.
2. Open the interface door.
3. Turn the extraction lens assembly counterclockwise, see
4. Pull out the extraction lens assembly, see
.
Figure 8-16.
Turn the extraction lens