Operation
Shutting Down the System
Thermo Fisher Scientific
iCAP Q Operating Manual (P/N 1288090, Revision B)
6-7
Shutting Down the System
Shutting down the system is only necessary for maintenance reasons.
Switching to Standby Mode
After the experiments are finished and the system was flushed, in the
evening or before weekends, the instrument can be set to standby
position.
Initial position:
•
The experiments are finished.
•
The plasma is still lit.
•
The system was flushed.
❖
To set the instrument to standby
1. Click
inside the Experiment Editor window:
•
The plasma is switched off.
•
Data that have not been saved before are saved now.
•
If installed, the autosampler moves to the home position.
•
The vacuum is still intact.
•
The instrument is in standby mode.
2. Release the tubings from the peristaltic pump to maintain their
flexibility.
Shutting Down the Complete System (Instrument Vented)
Initial position:
•
The experiments are finished.
•
The plasma is still lit.
•
The system was flushed.
NOTICE
If the plasma is not lit, start with
.
▲
❖
To shut down the instrument
1. Click
inside the Experiment Editor window. The
plasma is switched off.