Maintenance
Maintaining the Inlet System
Thermo Fisher Scientific
iCAP Q Operating Manual (P/N 1288090, Revision B)
8-13
3. Replace the sample uptake tubing. If it proves difficult to insert the
PTFE tubing in the ends, widen them by inserting the tip of a clean
micro pipette tip.
4. Replace the drain tubing.
5. Place the two tubes around the barrel rollers:
a. Make sure that they are not twisted.
b. Secure them in place using the collars and brackets.
6. Attach the tension arms.
7. Ensure that the tubes are fitted for correct sample and drain flow
direction when the peristaltic pump is in operation.
8. Switch on the peristaltic pump in the Instrument Control.
NOTICE
You must be logged in as Qtegra Administrator to access the
Instrument Control window.
▲
9. Check that the sample is being taken up into the instrument and
being drained efficiently.
10. If needed, adjust the tension of the tubing using the thumbscrews
on the peristaltic pump.
NOTICE
If an autosampler is being used, it is recommended to renew
the wash station tubing whenever the sample uptake and drain tubing
are changed.
▲
Maintaining the Spray Chamber
Removing the Spray Chamber
❖
To remove the spray chamber
1. Remove the nebulizer, see
.
2. Remove the Peltier cooling:
a. Remove the thumb screw at the Peltier cooling.
b. Carefully remove the lid.
c. Pull back the holder spring to release the spray chamber.
3. Carefully pull out the spray chamber and the elbow connector from
the torch assembly by pulling out the elbow connector.