Maintenance
Maintaining the Interface
8-24
iCAP Q Operating Manual (P/N 1288090, Revision B)
Thermo Fisher Scientific
5. Dismount the extraction lens assembly.
6. Check the extraction lens parts for damage or dirt:
•
Either, clean the extraction lens, see
•
Or, replace the extraction lens.
Cleaning the Extraction Lens
Figure 8-17.
Pull out the extraction lens
Figure 8-18.
Extraction lens
CAUTION
!
Corrosive Acids.
When handling the corrosive acids required to
perform the cleaning procedures described below, take the following
precautions.
•
Wear protective clothing.
•
Use acid resistant laboratory gloves and glasses.
•
Make yourself aware of all safety procedures in case of spillage or
exposure to skin or eyes before commencing the cleaning.
The acid cleaning should take place in a fume hood.